Certified Reference Material Management

Reference Material Management

The management of reference materials and certified solutions must be clearly organized. The conformity assessment body (CAB) must provide evidence of their metrological traceability to the SI.

Definitions

Definitions Related to Metrological Traceability

The laboratory must demonstrate that its equipment is calibrated using standards or devices whose measurement results are themselves traceable to the SI through a documented chain of calibrations performed by technically competent providers.

6.5 - Metrological Traceability

6.5.1 The laboratory shall establish and maintain metrological traceability of its measurement results by means of a documented, unbroken chain of calibrations, each contributing to the measurement uncertainty, linking them to an appropriate reference.

6.5.2 The laboratory shall ensure that measurement results are traceable to the International System of Units (SI) through:
a) calibration provided by a competent laboratory; or
b) certified values of certified reference materials provided by a competent producer, with stated metrological traceability to the SI; or
c) direct realization of SI units ensured by comparison, directly or indirectly, with national or international standards.

6.5.3 When metrological traceability to SI units is not technically possible, the laboratory shall demonstrate metrological traceability to an appropriate reference, such as:
a) certified values of certified reference materials supplied by a competent producer; or
b) results from reference measurement procedures, specified methods, or consensus standards that are clearly described and accepted by an appropriate authoritative body as providing measurement results fit for their intended use and ensured by suitable comparison.

Methodology

The conformity assessment body (CAB) must first identify the critical equipment used in its laboratory:

- One approach is to identify equipment that produces results directly reported to the customer and significantly affects the accuracy of an analysis, test, or calibration result, and therefore may affect the customer’s conclusions or statement of conformity.

Examples: a spectrometer used for chemical analysis (critical); a tensile testing machine used to determine tensile properties (critical); a chromatographic system used to detect a contaminant in an organic matrix (critical).

- A second approach is to assess whether the uncertainty contribution associated with equipment calibration materially affects the combined uncertainty of the final test result. If it is negligible, the equipment may be classified as noncritical for that measurement.

Examples: a room thermometer with an uncertainty of ±2 °C used to monitor ambient temperature where a tensile testing standard allows a range of 15 °C to 35 °C (noncritical in this example); a balance used to weigh a test portion for a chemical determination, whose uncertainty contributes directly to the final concentration uncertainty (critical).

For qualitative tests, the CAB assesses equipment criticality on the basis of the applicable standards, guidance documents, and experience.

Example: bending a specimen to check for possible surface cracks (the bending machine may be noncritical for this particular qualitative determination).

The key point is that standards, devices, solutions, and certified reference materials used to calibrate or verify critical equipment must be appropriately controlled. The laboratory must have evidence of the necessary metrological traceability and of the competence of the provider (for example, through accreditation to ISO/IEC 17025 or ISO/IEC 17034, where applicable).

The Metrological Traceability Hierarchy

Equipment and instruments used for testing and calibration, including accessories that can affect the validity of results (for example, environmental monitoring instruments), must be calibrated where required before being placed in service and subsequently at appropriate intervals. The purpose is to establish their measurement characteristics and associated uncertainty and to support metrological confirmation against applicable requirements.

To this end, the CAB should implement an appropriate calibration program ensuring metrological traceability to suitable references, ordinarily SI realizations, for measurement equipment whose performance affects the validity of results, except where:

- It has been demonstrated that the instrument’s contribution to the overall uncertainty is negligible. In such cases, the calibration and traceability needs must still be determined from the intended use and applicable requirements (for example, a room-temperature thermometer used during tensile testing of metallic materials).

- SI traceability is not technically achievable or does not apply to the measurand. In that case, confidence in the results may be established by other suitable means, such as:

• Use of reference materials, including certified reference materials, whose relevant chemical or physical properties have been adequately characterized;

• Use of clearly defined and widely accepted standards and measurement methods;

• Participation in appropriate comparisons, including interlaboratory comparisons.

The laboratory’s own reference standards (internal standards) must be calibrated so that their metrological traceability to appropriate references is established.

These reference standards should normally be reserved for calibration activities unless it is demonstrated that other uses do not compromise their performance.

Handling, transportation, and storage of standards and reference materials must follow appropriate procedures to preserve their condition and integrity.

Reference materials prepared within the laboratory must be produced using well-defined procedures that provide assurance of the required properties, where appropriate by using certified or equivalent materials.

Checks necessary to maintain confidence in the laboratory’s reference standards and reference materials must be performed in accordance with defined procedures and schedules where required.

Additional requirements may be necessary when such standards or materials are used away from the laboratory’s permanent premises.

Routes for Establishing Metrological Traceability to the SI:

The CAB using the measuring equipment must define its specifications and requirements based on the relevant analytical, testing, or calibration method and on the accuracy required for the results. These specifications inform the calibration program and the metrological confirmation process.

For critical measuring equipment, the internationally recognized traceability policy distinguishes the following routes. The designations below follow the framework of ILAC P10, scheduled to be replaced by Global ACI-TECH-1-011 (M). The successor reference should not be treated as effective until officially published and adopted.

Route 1: Calibration by a national metrology institute (NMI) or designated institute where the service is covered by the CIPM Mutual Recognition Arrangement (CIPM MRA). Coverage is established by the relevant calibration and measurement capability (CMC) published in the BIPM Key Comparison Database (KCDB).

The supporting evidence is the corresponding calibration report or certificate, identifying the results and associated uncertainty and permitting verification of the recognized service.

Route 2: Calibration by an accredited calibration laboratory where the specific service falls within the laboratory’s scope of accreditation and the accreditation body is a signatory to the applicable international or regional mutual recognition arrangement. Evidence is ordinarily an appropriately endorsed calibration certificate and the published accreditation scope.

Route 3a: Calibration by an NMI or designated institute for a service that is not covered by the CIPM MRA. The laboratory must obtain additional evidence supporting the provider’s technical competence and the claimed traceability.

Route 3b: Calibration by a calibration laboratory for a service that is not covered by the applicable international or regional accreditation recognition arrangements. Additional evidence is required to establish competence, uncertainty evaluation, and metrological traceability.

Route 3b can include, depending on the recognition status and scope of the service, the following arrangements:

3b1) In-house calibration: calibration performed by the testing or calibration laboratory for its own use. The laboratory retains an internal calibration record or certificate documenting the method, measurement results, uncertainty, applicable standards, and traceability.

When an accredited laboratory, or an applicant for accreditation, establishes internal calibration arrangements for critical measuring equipment using its own reference standards, its technical competence should be supported by appropriate evidence, including:
- Verification or validation of the calibration method;
- Evaluation of measurement uncertainty;
- Metrological traceability of reference standards and measurement results;
- Monitoring the validity of calibration results;
- Personnel competence;
- Facilities and environmental conditions;
- Internal audits covering the calibration activities.

3b2) Internal metrology service: calibration performed by a centralized or decentralized metrology service belonging to the same company or corporate group as the accredited laboratory. The laboratory must determine whether that service is covered by an appropriate accreditation and, if not, establish and document the relevant evidence of its technical competence.

For the relevant quantities and calibration ranges, the internal metrology service’s ability to produce technically valid calibration results must be assessed based on the applicable requirements of ISO/IEC 17025, including method suitability, uncertainty, traceability, personnel competence, equipment, and records. An internal organizational relationship alone does not demonstrate competence or international recognition.

Routes 3a and 3b may be used when routes 1 and 2 are not available for the required service, provided that the applicable accreditation policy permits them and the necessary technical evidence is available. Whenever feasible, a recognized calibration service under route 1 or 2 remains the preferred solution. Note: in-house calibration is not automatically “route 3b”; classification depends on the recognition status of the calibration service.

For additional internationally agreed principles, see the BIPM–OIML–ILAC–ISO Joint Declaration on Metrological Traceability. This declaration provides common principles rather than detailed rules for assessing non-recognized calibration services.