Gage Repeatability and Reproducibility Studies: Range Method

Average and Range Method:

The Gage R&R method, which stands for Repeatability and Reproducibility, calculates the total variability of a measurement system and makes it possible to separate repeatability, reproducibility, and part-to-part variation.

To quantify repeatability and reproducibility using the average and range method, several parts, one measuring instrument, several operators, and several measurements are required.

The recommended method is to use 10 parts, 3 operators, and 3 measurement trials, for a total of 90 measurements.

La répétabilité EV du système de mesure est donnée par la formule:

is the average of all the ranges obtained by the three operators across all parts.

d2 is the constant selected from the table below , where Z is obtained by multiplying the number of parts n by the number of operators (Z = 10 × 3 = 30), and w is the number of trials (w = 3).

5,15 corresponds to the number of standard deviations required to obtain a 99% coverage probability.

The measurement system’s AV reproducibility is given by the following formula:

is the average difference between the mean measurements obtained by the operator with the highest measurements and the operator with the lowest measurements, across all operators and all parts.

n = 10: Number of parts
r = 3: Number of repeated trials

d2 is the constant selected from the table below with Z = 1 et w is the number of operators (w=3).

Repeatability and reproducibility are calculated using the following formula:

The variability associated with the part is given by the formula:

Rp : For each operator, it corresponds to the difference between the average measurement of the largest part and that of the smallest part..

d2 is the constant selected from the table below with Z = 1 et w is the number de parts (w=10).

Total variability is obtained by combining the measurement system variability and the part variability in quadrature:

Numerical example:

Three operators, OP1, OP2, and OP3, measure a 45 mm dimension L using a calibrated traceable caliper. The tolerance for the dimension is ±1 mm.

Measurement number

OP1

x̄
R1
OP2
x̄
R2
OP3
x̄
R3
R̄
1
45,08
45,10
45,09
45,09
0,02
45,01
45,04
45,02
45,02
0,03
45,04
45,07
45,07
45,06
0,03
45,06
0,027
2
44,94
44,97
44,93
44,95
0,04
44,92
44,95
44,95
44,94
0,03
44,92
44,97
44,91
44,93
0,06
44,94
0,043
3
45,08
45,11
45,13
45,11
0,05
44,95
45,11
45,11
45,06
0,16
45,10
45,12
45,11
45,11
0,02
45,09
0,077
4
44,97
45,02
45,01
45,00
0,05
45,01
44,93
45,00
44,98
0,08
45,02
44,99
44,99
45,00
0,03
44,99
0,053
5
44,92
44,95
45,01
44,96
0,09
44,95
44,99
44,97
44,97
0,04
44,99
44,97
44,95
44,97
0,04
44,97
0,057
6
45,03
45,06
45,05
45,05
0,03
45,04
45,06
44,99
45,03
0,07
45,07
45,03
45,02
45,04
0,05
45,04
0,050
7
45,24
45,36
45,31
45,30
0,12
45,12
45,21
45,32
45,22
0,20
45,26
45,29
45,28
45,28
0,03
45,27
0,117
8
45,16
45,16
45,10
45,14
0,06
45,12
45,13
45,13
45,13
0,01
45,13
45,13
45,14
45,13
0,01
45,13
0,027
9
45,20
45,20
45,24
45,21
0,04
45,06
45,19
45,24
45,16
0,18
45,19
45,17
45,16
45,17
0,03
45,18
0,083
10
45,07
45,10
45,05
45,07
0,05
45,02
45,05
45,09
45,05
0,07
45,10
45,10
45,09
45,10
0,01
45,07
0,043
45,09 
0,055
 
 
 
45,06
0,087
45,08
0,031
R̄1
R̄2
R̄3
  Moyenne générale
 45,07
0,058

Let’s calculate the repeatability of the measurement system:

=0,058

d2=1,693

EV=Repeatability = 0,176

Let’s calculate the reproducibility of the measurement system:

=difference between OP1 (highest readings) and OP2 (lowest readings)=0,032

d2=1,91; n=10 et r=3

AV=Reproducibility = 0,080

Calculons le R&R:

EV=Repeability= 0,176 et AV=Reproducibility = 0,080

R&R = √(EV²+AV²)=0,193

Let’s calculate the part-to-part variability.:

Rp (Range of the means of means) = 45,27-44,94 = 0,33

d2=3,18

VP=0,53

Let’s perform the Gage R&R calculation:

VT = √(R&R²+VP²)=0,56

We then determine the percentage contributions :

%EV= (EV/TV).100 = (0,176/0,56).100 = 31,4 % (Equipment)

%AV=(AV/TV).100 = (0,08/0,56).100 = 14,3 % (Operator)

%R&R = (R&R/VT).100 = (0,193/0,56).100 = 34,4%

%VP = (PV/TV).100 = (0,534/0,564).100 = 94,6 % (Part)

If the R&R percentage is:
< 10%: the process is satisfactory.
Between 10% and 30%: the process is acceptable, but could be improved.
> 30%: the process is unacceptable, which is the case here.
In this case, the process is degraded by operator repeatability, which is not optimal.

This may be explained by an instrument that is not well suited to the type of dimension being measured, or by a measurement method that does not allow for repeatable results, such as a complex-shaped area or imprecise positioning of the instrument.

According to MSA guidance, when a Gage R&R study is conducted with only one piece of equipment, reproducibility should normally be forced to zero, and AV should not be included in the analysis.

If the analysis is focused on the Product, the Tolerance is used instead of the Total Variation.

Both approaches, Total Variation and Tolerance, are valid. The choice depends on how the measurement system is used and on the customer’s requirements.

When the analysis is based on Total Variation, it must be ensured that the measured parts properly represent the Total Process Variation.

Tolérance = IT = 2

%TOL EV = (0,176/2).100 = 8,8%

%TOL AV = (0,08/2).100 = 4,0%

%R&R = (0,193/2).100 = 9,6%

%PV = (0,534/2).100 = 26,7%

In the present case, the product analysis is satisfactory.

Table des valeurs de d2