Gage Repeatability and Reproducibility Studies: Range Method
Average and Range Method:
The Gage R&R method, which stands for Repeatability and Reproducibility, calculates the total variability of a measurement system and makes it possible to separate repeatability, reproducibility, and part-to-part variation.
To quantify repeatability and reproducibility using the average and range method, several parts, one measuring instrument, several operators, and several measurements are required.
The recommended method is to use 10 parts, 3 operators, and 3 measurement trials, for a total of 90 measurements.
La répétabilité EV du système de mesure est donnée par la formule:

d
5,15 corresponds to the number of standard deviations required to obtain a 99% coverage probability.
The measurement system’s AV reproducibility is given by the following formula:

is
the average difference between the mean measurements obtained by the operator
with the highest measurements and the operator with the lowest measurements,
across all operators and all parts.
n = 10: Number of parts
r = 3: Number of repeated trials
d
Repeatability and reproducibility are calculated using the following formula:

The variability associated with the part is given by the formula:

R
d
Total variability is obtained by combining the measurement system variability and the part variability in quadrature:
![]()
![]()
Three operators, OP1, OP2, and OP3, measure a 45 mm dimension L using a calibrated traceable caliper. The tolerance for the dimension is ±1 mm.
| Measurement number |
OP1 |
x̄
|
R1 |
OP2 |
x̄ |
R2 |
OP3 |
x̄ |
R3 |
|
R̄ |
||||||
| 1 |
45,08 |
45,10 |
45,09 |
45,09 |
0,02 |
45,01 |
45,04 |
45,02 |
45,02 |
0,03 |
45,04 |
45,07 |
45,07 |
45,06 |
0,03 |
45,06 |
0,027 |
| 2 |
44,94 |
44,97 |
44,93 |
44,95 |
0,04 |
44,92 |
44,95 |
44,95 |
44,94 |
0,03 |
44,92 |
44,97 |
44,91 |
44,93 |
0,06 |
44,94 |
0,043 |
| 3 |
45,08 |
45,11 |
45,13 |
45,11 |
0,05 |
44,95 |
45,11 |
45,11 |
45,06 |
0,16 |
45,10 |
45,12 |
45,11 |
45,11 |
0,02 |
45,09 |
0,077 |
| 4 |
44,97 |
45,02 |
45,01 |
45,00 |
0,05 |
45,01 |
44,93 |
45,00 |
44,98 |
0,08 |
45,02 |
44,99 |
44,99 |
45,00 |
0,03 |
44,99 |
0,053 |
| 5 |
44,92 |
44,95 |
45,01 |
44,96 |
0,09 |
44,95 |
44,99 |
44,97 |
44,97 |
0,04 |
44,99 |
44,97 |
44,95 |
44,97 |
0,04 |
44,97 |
0,057 |
| 6 |
45,03 |
45,06 |
45,05 |
45,05 |
0,03 |
45,04 |
45,06 |
44,99 |
45,03 |
0,07 |
45,07 |
45,03 |
45,02 |
45,04 |
0,05 |
45,04 |
0,050 |
| 7 |
45,24 |
45,36 |
45,31 |
45,30 |
0,12 |
45,12 |
45,21 |
45,32 |
45,22 |
0,20 |
45,26 |
45,29 |
45,28 |
45,28 |
0,03 |
45,27 |
0,117 |
| 8 |
45,16 |
45,16 |
45,10 |
45,14 |
0,06 |
45,12 |
45,13 |
45,13 |
45,13 |
0,01 |
45,13 |
45,13 |
45,14 |
45,13 |
0,01 |
45,13 |
0,027 |
| 9 |
45,20 |
45,20 |
45,24 |
45,21 |
0,04 |
45,06 |
45,19 |
45,24 |
45,16 |
0,18 |
45,19 |
45,17 |
45,16 |
45,17 |
0,03 |
45,18 |
0,083 |
| 10 |
45,07 |
45,10 |
45,05 |
45,07 |
0,05 |
45,02 |
45,05 |
45,09 |
45,05 |
0,07 |
45,10 |
45,10 |
45,09 |
45,10 |
0,01 |
45,07 |
0,043 |
| 45,09 |
0,055 |
|
45,06 |
0,087 |
45,08 |
0,031 |
|||||||||||
| |
R̄1 |
|
R̄2 |
|
R̄3 |
||||||||||||
|
Moyenne générale |
|
45,07 |
|
0,058
|
|||||||||||||
Let’s calculate the repeatability of the measurement system:
=0,058
d
EV=Repeatability = 0,176
Let’s calculate the reproducibility of the measurement system:
![]()
=difference between OP1 (highest readings) and OP2 (lowest readings)=0,032
d
AV=Reproducibility = 0,080
Calculons le R&R:
EV=Repeability= 0,176 et AV=Reproducibility = 0,080
R&R = √(EV²+AV²)=0,193
Let’s calculate the part-to-part variability.:
R
d
VP=0,53
Let’s perform the Gage R&R calculation:
VT = √(R&R²+VP²)=0,56
![]()
We then determine the percentage contributions :
%EV= (EV/TV).100 = (0,176/0,56).100 = 31,4 % (Equipment)
%AV=(AV/TV).100 = (0,08/0,56).100 = 14,3 % (Operator)
%R&R = (R&R/VT).100 = (0,193/0,56).100 = 34,4%
%VP = (PV/TV).100 = (0,534/0,564).100 = 94,6 % (Part)
If the R&R percentage is:
< 10%: the process is satisfactory.
Between 10% and 30%: the process is acceptable, but could
be improved.
> 30%: the process is unacceptable, which is the case
here.
In this case, the process is degraded by operator repeatability, which is
not optimal.
This may be explained by an instrument that is not well suited to the type of dimension being measured, or by a measurement method that does not allow for repeatable results, such as a complex-shaped area or imprecise positioning of the instrument.
If the analysis is focused on the Product, the Tolerance is used instead of the Total Variation.
Both approaches, Total Variation and Tolerance, are valid. The choice depends on how the measurement system is used and on the customer’s requirements.
When the analysis is based on Total Variation, it must be ensured that the measured parts properly represent the Total Process Variation.
Tolérance = IT = 2
%TOL EV = (0,176/2).100 = 8,8%
%TOL AV = (0,08/2).100 = 4,0%
%R&R = (0,193/2).100 = 9,6%
%PV = (0,534/2).100 = 26,7%
In the present case, the product analysis is satisfactory.